Télécharger Imprimer la page

FLAEM RF6 Basic 2 Mode D'emploi page 12

Masquer les pouces Voir aussi pour RF6 Basic 2:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
low.
method A: Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6 can be dis-
infected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlorite) specific for disinfect-
ing, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable water and disinfectant,
according to the proportions indicated on the packaging of the disinfectant.
- Completely immerse each part in the solution, taking care to avoid the formation of air bubbles on the parts.
Leave the parts immersed for the amount of time indicated on the packaging of the disinfectant associated with
the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
- Dispose of the solution following the instructions provided by the disinfectant manufacturer.
method B: Disinfect the accessories C1.1-C1.2-C1.3-C1.4-C2-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6 by
boiling them in water for 10 minutes; use demineralised or distilled water to prevent calcium deposits.
method C: Disinfect the accessories 1C1.1-C1.2-C1.3-C1.4-C2-C5.1A-C5.1B-C5.1C-C5.2-C5.3-C5.4-C5.5-C6
with a hot steam steriliser for baby-bottle (not the microwave type). Perform the process faithfully following the
instructions of the steriliser. To ensure that the disinfection is effective, choose a steriliser with an operating cycle
of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a paper towel. Alternatively,
dry them with a jet of hot air (for example, from a hair dryer).
At the end of each use store the device complete with accessories in a dry place away from dust.
AIR FILTERING
The unit is equipped with an air filter (A3) which should be replaced when dirty or when its
colour changes. Do not wash or re-use the used filter. The filter must be regularly replaced
to help ensure efficient compressor performance. The filter should be regularly checked.
For spare filters contact your dealer or an authorised service centre.
To replace the filter:
Lift the lid and remove the filter using a screwdriver (see fig.). The filter is designed
to remain fixed in its housing.
Only use original accessories and spare parts by Flaem, we disclaim any liability in
the event of using non original spare parts or accessories.
10
(A3)

Publicité

loading

Ce manuel est également adapté pour:

Rhino clearLella la coccinella