6 Maintenance
6.3 Color calibration
78
Color-calibrating the scanner
NOTE
Only use color calibration set with clean, dry CEREC Primescan /
Primescan / CEREC Omnicam / Omnicam scanner
In order to achieve optimum results, the CEREC Primescan /
Primescan / CEREC Omnicam / Omnicam scanner must be clean,
disinfected and dry before color calibration.
➢ Make sure that the CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam scanner is clean, disinfected and dry.
1. Remove the color-calibration set from the packaging.
2. Use the CEREC Primescan / Primescan / CEREC Omnicam /
Omnicam scanner to scan the QR code on the underside of your
color-calibration set. In order to do this, you must hold the CEREC
Primescan / Primescan / CEREC Omnicam / Omnicam scanner still
in front of the QR code so that it is completely visible in the picture.
If the QR code appears to be shiny, hold the scanner at more of an
oblique angle in order to avoid any glaring light and to make it
easier to scan the codes. If the QR code is recognized, the next
"Please mount color calibration set" step appears.
This step of the QR code scan is skipped during the subsequent
color calibration and the serial number of the color-calibration set is
thus displayed. If this does not match the serial number printed on
your color-calibration set, click on the "Rescan QR Code" button
and scan the new QR code.
3. Mount the color-calibration set on the tip of the scanner until it locks
into place.
4. Click on the "Ok" button.
Ä The measuring process starts. Do not move the CEREC
Primescan / Primescan / CEREC Omnicam / Omnicam scanner
or the color-calibration set during this time.
Ä The software provides status updates on the calibration and
informs you once the procedure is complete.
Ending the color calibration
ü The software indicates that the color calibration was completed
successfully.
1. Click on the "Ok" button.
Ä The CEREC Primescan / Primescan / CEREC Omnicam /
Omnicam scanner is now color-calibrated.
2. Remove the color-calibration set from the scanner and place it back
in the packaging.
Error message during color calibration
The software indicates if an error occurs during color calibration. If the
color calibration contained an error, ensure the following:
● The color-calibration set is free of dust
● The color-calibration set was mounted correctly
● The CEREC Primescan / Primescan / CEREC Omnicam / Omnicam
scanner exit window is clean
➢ Then restart the color calibration.