Calibrating The Scanner - Dentsply Sirona CEREC Primescan AC Notice D'utilisation

Masquer les pouces Voir aussi pour CEREC Primescan AC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6 Maintenance

6.2 Calibrating the scanner

74
Calibrating the scanner
6.2
The measurement procedure used by the system requires the use of a
calibrated scanner. The CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam scanner is calibrated ex works. Calibrate the
scanner after every reinstallation and after each time that it is
transported. The calibration set supplied is available for the calibration
process.
In order to achieve optimum results, the CEREC Primescan /
Primescan / CEREC Omnicam / Omnicam scanner must be allowed to
warm up for 2 minutes before calibration.
Recalibrate the scanner in the following cases:
● following transport (shaking stress) or during first commissioning,
● after storage in unheated or un-air-conditioned rooms (temperature
differences exceeding 30°C / 85°F),
● with temperature differences of over 15°C / 60°F between the last
calibration and operation.
● In general, carrying out a calibration is the correct process in the
event of errors in the acquisition process (such as poor image
quality or the lack of a 3D preview). In many cases, the errors can
be corrected in doing so.
● As the system may be exposed to vibration loads without knowledge
of this, it should be calibrated once a month.
Starting calibration
1. In the software, navigate to the system menu and click on the
"Configuration" button.
2. Click on the "Devices" button.
3. Click on the "Primescan" button.
4. Click on the "Calibrate" button.
Ä The scanner view is displayed in one window.
Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02    04.2019

Hide quick links:

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Primescan acCerec omnicam acCerec primescan aOmnicam ac

Table des Matières