Dentsply Sirona CEREC Primescan AC Notice D'utilisation page 37

Masquer les pouces Voir aussi pour CEREC Primescan AC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Dentsply Sirona
Operating Instructions
Interference immu-
IEC 60601-1-2 Test level
nity tests
Conducted RF in-
3 V
terference
150 kHz to 80 MHz
IEC 61000-4-6
6  V
in ISM frequency bands
between 150  k Hz and 80
MHz
80% AM at 1  k Hz
Radiated RF inter-
3  V /m
ference
80 MHz to 800  M Hz
IEC 61000-4-3
3  V /m
800  M Hz to 2.7  G Hz
80% AM at 1  k Hz
Immunity to interference against high-frequency electromagnetic fields in the direct vicinity of wireless communi-
cation devices IEC 61000-4-3
Test frequency
Modulation
(MHz)
385
Pulse modulation: 18   H z
450
FM + 5  H z difference: 1
  k Hz Sinus
710
Pulse modulation: 217
  H z
745
780
810
Pulse modulation: 18   H z
870
930
1720
Pulse modulation: 217
  H z
1845
1970
67 15 481 D3696
D3696.201.01.03.02    04.2019
Compliance level
3  V
eff
6  V
eff
3  V /m
3  V /m
Required immunity test
level (V/m)
27
28
9
28
28
Electromagnetic environment -
guidelines
d= [1.2] √P
eff
eff
d= [1.2] √P
at 80 MHz to 800 MHz
d= [2.3] √P
at 800 MHz to 2.7 GHz
with P as the power rating of the
transmitter in watts (W) according to
the transmitter manufacturer's speci-
fications and d as recommended
safety distance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey
than the compliance level
frequency range.
Interference is possible in the vicinity
of equipment bearing the following
graphic symbol.
Maintained immunity test level (V/m)
27
28
9
28
28
3 Product description
3.6 Electromagnetic compatibility
1
should be less
2
in each
33

Hide quick links:

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Primescan acCerec omnicam acCerec primescan aOmnicam ac

Table des Matières