Operating Instructions
Note on wireless communication
67 15 481 D3696
D3696.201.01.03.02    04.2019
2

Safety instructions

2.1

Disturbance of data transmission

Data communication between the acquisition unit and the CEREC MC
XL production unit should preferably be established via the wireless
interface CEREC Radio Device or WLAN.
As for all wireless connections (e.g. cell phones), heavy utilization of the
available radio channels or shielding caused by building installations
(e.g. metal-shielded X-ray enclosures) may impair the quality of the
connection. This may become noticeable through a reduction in range
and/or a slower data transmission rate. In extreme cases, it will be
impossible to establish a wireless connection at all.
Dentsply Sirona has selected the best possible configuration for data
communication via the wireless interface (CEREC Radio Device) or
WLAN, which generally ensures perfect functioning of this connection.
However, in individual cases unrestricted wireless data communication
may be impossible for the reasons mentioned above and/or due to local
circumstances. In such cases, a cable LAN connection should be
selected to ensure uninterrupted operation. If the only LAN interface on
the rear of the acquisition unit is occupied by another plug, remove this
wireless interface connection and instead connect the LAN cable with
the CEREC MC XL production unit.

Basic safety information

2.2
2.2.1

Prerequisites

NOTE
Important information on building installation
In order to prevent the risk of an electric shock, this unit must only be
connected to a supply mains with a ground wire.
The building installation must be performed by a qualified expert in
compliance with the national regulations.
NOTE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
2.2.2

Connecting the unit

Perform connection by following the directions given in the present
operating instructions.
2 Safety instructions