11. Remove the lid and the holder with the sleeve from the container.
12. Thoroughly rinse the sleeve with tap water of potable water quality
that meets Federal Clean Water Standards holding it with its holder
(for at least 30 seconds).
13. Carefully and slowly remove the plug with the holder from the
sleeve by downward movement of the holder. Make sure that the
sleeve points upwards not to risk any damage by liquids entering
the inside of the sleeve.
14. Remove the plug from the holder.
15. Dry the plug with a clean lint-free cloth.
16. Prior to each use of the HLD set test the solution for its
effectiveness per manufacturer´s instructions. If not in use rinse
container and holder and store the holder inside the empty
container.
6 Maintenance
6.1 Cleaning, disinfection and sterilization
69