6 Maintenance
6.1 Cleaning, disinfection and sterilization
62
6.1.4.1

General information

The CEREC Primescan / Primescan / CEREC Omnicam / Omnicam is a
very sensitive optical device and must therefore be handled with the
utmost care. Protect the coated sapphire glass and the camera window
against scratches and clean them with a lint-free cloth and ethanol
(commercially available cleaning alcohol) if any haze is noted during the
acquisition. Wipe down the window afterwards again with the absorbent
cotton gauze dipped in drinking water.
NOTE
Hygiene processes
Observe the following hygiene processes.
Dental health care providers are advised to select the reprocessing
method that aligns with their infection control process. An effective
infection control program is practical, reasonable, and reproducible.
The following methods have been validated to reprocess the CEREC
Primescan / Primescan / CEREC Omnicam / Omnicam between patient
care:
● High-level disinfection of the sleeve (with HLD Set)  [→ 66]
● Dry heat sterilization of the sleeve  [→ 71]
● Using disposable sleeves  [→ 72]
One of these methods is strongly recommended to be used for the
reprocessing of the CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam.
Dentsply Sirona
67 15 481 D3696
D3696.201.01.03.02    04.2019