2 Safety instructions
2.3 Safety labels
18
Plug connections of external interfaces
CAUTION
Adaptation of acquisition unit to external components
Additional devices connected to external interfaces must be tested
according to the relevant standards, e.g.:
EN 60601-1:2006 + Cor.:2010 + A1:2013,
IEC 60601-1 Edition 3.1:2012,
EN 61010-1:2010 based on IEC 61010-1:2010 + Cor.:2011.
They must be installed outside of the patient area (a radius of 1.5m
surrounding the patient).
WARNING
Risk of electric shock
Low voltages are applied to the sockets for connecting external
interfaces. In order to maintain electrical safety, the rear covers of the
acquisition unit must be kept closed while it is in operation (service
cover and cover on the monitor).
➢ Do not touch the pins of the connectors.
➢ When using the unit on the patient, please note that the covers on
the rear side of the unit (service cover and cover on the monitor)
must remain closed and voltage sources must not be accessible.
The cover on the monitor must not be opened, if both USB
sockets are in use or locked.
➢ The acquisition unit must not be operated inside of the patient
area (within a radius of 1.5 m surrounding the patient) with the
covers open.
NOTE
Risk of damage to the plugs/cables!
The externally connected plugs/cables may be damaged, if they are
overtensioned or if the plug connections do not snap in.
➢ Do not pull on the cables.
➢ Make sure that the plug connections snap in.