4 Installation and startup
4.4 Initial startup
44
NOTE
Reduced battery service life
If the battery is not charged over a long period of time, this
significantly reduces its service life.
➢ Always recharge the battery fully after buffer operation.
NOTE
In battery mode, the CEREC Primescan AC / Primescan AC / CEREC
Omnicam AC / Omnicam AC acquisition unit must remain connected
to the mains supply for at least two hours after using the battery
buffering to charge the storage battery.
At least every six weeks the CEREC Primescan AC / Primescan AC /
CEREC Omnicam AC / Omnicam AC acquisition unit should be
connected to the mains supply to charge the storage batteries.
IMPORTANT
The storage battery is only charged if the capacity falls below the
threshold of 90%.
In the case of ambient temperatures above 28°C and unfavorable
operating conditions, this can result in delays when charging the
battery.
Dentsply Sirona
Operating Instructions
67 15 481 D3696
D3696.201.01.03.02 04.2019