12. Technical specifications

Model no.
BM 26
Measurement
Oscillometric, non-invasive blood pressure
method
measurement on the upper arm
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 60 – 280 mmHg,
diastolic 30 – 200 mmHg,
pulse 30 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5% of the value shown
Measurement
Max. permissible standard deviation
uncertainty
according to clinical testing:
systolic 8 mmHg /
diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
L 155 mm x W 110 mm x H 70 mm
Weight
Approximately 395 g
(without batteries, with cuff)
Cuff size
22 to 35 cm
Permissible
+10°C to +40°C, 15-90% relative humidity
operating
(non-condensing)
conditions
Permissible
- 25°C to + 70°C, ≤ 93% relative humidity,
storage conditions
700 – 1060 hPa ambient pressure
Power supply
4 x 1.5 V
type LR6)
AA batteries (alkaline
Battery life
For approx. 180 measurements, depending
on levels of blood pressure and inflation
pressure
Classification
Internal supply, IPX0, no AP or APG, con-
tinuous operation, application part type BF
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This device conforms with the European standard
EN 60601-1-2 (in accordance with CISPR 11, IEC 61000-3-2,
IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6,
IEC 61000-4-8, IEC 61000-4-11) and is subject to particu-
lar precautions with regard to electromagnetic compatibili-
ty. Please note that portable and mobile HF communication
systems may interfere with this device.
• This device complies with the EU Medical Devices Directive
93/42/EEC, the German Medical Devices Act (Medizinpro-
duktgesetz) and the standards EN 1060-3 (Non-invasive
sphygmomanometers – Part 3: Supplementary requirements
for electro-mechanical blood pressure measuring systems)
and IEC 80601-2-30 (Medical electrical equipment – Part 2 –
30: Particular requirements for the basic safety and essential
performance of automated non-invasive sphygmomanom-
eters).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If the device is used for commercial medical purposes, it