Factory default settings

7
Factory default settings
7.1

Settings which can be changed via the menu

Menu
Relay active on alarm / No alarm
A1 Alarm
A2 Alarm
A1 alarm: direction
(falling for DrägerSensor O2)
A2 alarm: direction
A1 latch mode
A2 latch mode
A1 acknowledgment mode
A2 acknowledgment mode
Normal operation A1 relay
Normal operation A2 relay
Hysteresis for A1 alarm at direction rising 0
Hysteresis for A2 alarm at direction rising 0
Calibration password
Settings password
LCD setting
SIL status
Language
Fault current
Warning
Warning current
Warning cycle time T1
Warning cycle time T2
Maintenance
Maintenance current
Analog offset
Analog span
HART address
Auto-calibration
Calibration interval
Function key
Display capture offset at O2 sensor
Display capture low at O2 sensor
Display capture high at O2 sensor
2)
Modbus address
2)
Modbus parity
2)
Modbus baudrate
2)
Profibus address
2)
Fieldbus interfaces are not in accordance with the BVS 13 ATEX G001 X. Using the fieldbus interface no relays are available.
36
Factory Default Setting
Active on alarm
Depending on the sensor
Depending on the sensor
Rising
Rising
Non-latching
Latching
Acknowledgeable
Not acknowledgeable
Energized
Energized
_ _ _1
_ _ _2
On
Off
EN
1.2 mA
Off
3.0 mA
10 sec
1 sec
static
3.4 mA
0 mA
0 mA
0
Off
Depending on sensor
Faults
20.9 Vol%
-0.25 Vol%
0.25 Vol%
126
even (default)
9600 bps
126
Range
On / Off
Falling / Rising
Latching / Non-Latching
Acknowledgeable / Not acknowledgeable
/ Pre-acknowledgeable
Energized / Not energized
0 to A1
0 to A2
On / Off
On / Off
DE / EN / FR / ES / RU / ZH
0 to 3.5 mA
On / Off
0 to 3.5 mA
5 to 60 sec.
1 to 59 sec.
Static / Dynamic
0 to 3.5 mA
-0.5 to 0.5 mA
-0.5 to 0.5 mA
0 to 15
On / Off
0 to 720
Graph, fault, warning, vitality, bump test
0 to 25 Vol% (EC O2 LS)
0 to 100 Vol% (EC O2)
0 to -0.75 Vol%
0 to 0.75 Vol%
2-126
even, odd parity, no parity
9600 or 19200 bps
2-126
Dräger Polytron 8100