Troubleshooting; Fault Reference - Dräger Polytron 8100 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 88
6.3

Troubleshooting

6.3.1

Fault reference

Fault number
Cause
#001
Serious instrument fault,
#003
various causes.
#004
#005
#011 - #014
#020 - #024
#043
#060
#067
#002
Serious data error in the instrument,
#025 - #027
various causes.
#030 - #034
#041
#042
#044
#050
#052
#081
#085
#010
4 to 20 mA interface cable not connected.
#045
Instrument not detecting any sensor.
#051
Faulty zero calibration.
#054
#055
Faulty span calibration.
#076
Instrument fault.
#080 - #084
#090
Sensor lock function is activated. A sensor
with a different part number has been
connected.
#105
Instrument fault.
#137
Instrument fault.
®
1
DrägerService
is a registered trademark of Dräger.
Dräger Polytron 8100
Remedy
Have the instrument checked by DrägerService
Reset the instrument to factory default settings (see Section 7
on Page 36).
If this fault occurs again:
Have the instrument checked by DrägerService.
Disconnect the power supply, connect the 4 to 20 mA interface
and reconnect the power supply.
If operated as a stand-alone instrument without a central
controller, wire pin 3 to pin 4 (see Section 3.5 on Page 10).
Check connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
Perform zero calibration (see Section 6.2.1 on Page 31).
Perform span calibration (see Section 6.2.2 on Page 31).
Check electrical connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
Use a sensor with the same part number or disable the sensor
lock function (see Section 5.10.3 on Page 27).
Have the instrument checked by DrägerService.
Cycle power.
If this fault occurs again:
Have the instrument checked by DrägerService.
Maintenance
®1
.
33