Zero Calibration; Span Calibration - Dräger Polytron 8100 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 88
6.2.1

Zero Calibration

WARNING
!
Ambient air can be used to zero the sensor instead of
Nitrogen or Synthetic Air only if the area is known to be
free of the target gas or any gas to which the sensor
may be cross-sensitive (as listed on the sensor data
sheet). In this case, no cylinder or calibration adapter
is needed for the zero calibration.
NOTICE
i
i
For Oxygen (O
) sensors a zero calibration does not
2
change any value in the firmware or sensor. Thus, the
zero calibration function can be used to check the
correct zero reading when Nitrogen is applied to the
sensor. The Polytron 8100 will indicate a fault state if a
zero check fails with a deviation of more than 0.6 Vol%
O
from zero. In case of a fault, repeat the zero check
2
or change the sensor if necessary. Use Nitrogen (N
for verification.
1. Select Calibration > Zero Calibration and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The message Apply zero gas is displayed.
2. Apply Synthetic Air or Nitrogen.
3. Select Next and confirm.
The current value is displayed.
After the displayed value is stable:
1. Select Calibrate and confirm.
The message Please wait... is displayed.
The new current value is displayed.
2. Select Next and confirm.
3. Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
4. Select Next and confirm.
The instrument returns to the calibration menu.
Dräger Polytron 8100
6.2.2

Span calibration

1. Select Calibration > Span calibration and confirm.
The Maintenance signal is transmitted and the symbol [ ]
is displayed.
The parameters for the calibration gas are displayed, e.g.:
Gas
H2S
Unit
ppm
Conc.
25
Depending on the sensor these parameters can be changed:
1. Select Gas and confirm.
2. Select the calibration gas from the list and confirm.
3. Select Unit and confirm.
4. Select the unit of measurement from the list and confirm.
5. Select Conc. and confirm.
6. Set the concentration of the calibration gas.
)
2
If the settings are correct:
1. Select Next and confirm.
A message like Gas flow ON H2S is displayed.
2. Apply calibration gas.
3. Select Next and confirm.
The current value is displayed.
After the displayed value is stable:
1. Select Next and confirm.
The message Please wait... is displayed.
The new current value is displayed.
2. Select Next and confirm.
3. Turn off gas flow and remove the calibration adapter from
the sensor or disconnect tubing.
If the current value is not within the alarm range:
4. Select Next and confirm.
The instrument returns to the calibration menu.
A calibration can be aborted at any time.