Technical Data - Dräger X-am 5600 Notice D'utilisation

Masquer les pouces Voir aussi pour X-am 5600:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
Care
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning
solvents can destroy the dust and water filters.
Carefully dab dry the instrument using a cloth.
Maintenance
The instrument should be inspected and maintained annually by suitably
qualified persons (consult: EN 60079-29-2 – Guide for the selection, installation,
use and maintenance of apparatus for the detection and measurement of
combustible gases or oxygen; EN 45544-4 – Electrical apparatus used for the
direct detection and direct concentration measurement of toxic gases and
vapours - Part 4: Guide for selection, installation, use and maintenance and
national regulations).
Recommended calibration interval for measuring channels Ex, O
CO
: 12 months.
2
NOTICE
Calibration interval of other gases: see Instructions for Use for the respective
DrägerSensors.
Disposing of the instrument
EC-wide regulations for the disposal of electric and electronic appliances
which have been defined in the EC Directive 2002/96/EC and in national
laws have been effective since August 2005 and apply to this instrument.
Common household appliances can be disposed of using special collecting and
recycling facilities. However, as this instrument has not been registered for
household usage, it must not be disposed of through these means. The
instrument can be returned to your national Dräger Sales Organisation for
disposal. Please do not hesitate to contact the above if you have any further
questions on this issue.
20
CAUTION
, H
S, CO and
2
2

Technical data

Extract: see the Technical Handbook for details
Ambient conditions:
during operation and
–20 to +50
storage
type 180AAHC)
700 to 1300 hPa
10 to 90 % (to 95 % briefly) r.h.
Protection class
IP 67 for instrument with sensors
Alarm volume
Typically 90 dB (A) in 30 cm distance
Operation time
– Alkaline battery/
Typically 9 hours under standard conditions
NiMH rechargeable
batteries
- NiMH power pack
Typically 9 hours under standard conditions
Dimensions
approx. 130 x 48 x 44 mm (H x W x D)
Weight
approx. 220 to 250 g
CE marking:
Electromagnetic compatibility (Directive 89/336/
EEC)
Low Voltage Directive
(Directive 72/23/EEC)
Explosion protection
(Directive 94/9/EEC)
Approvals:
(see"Notes on Approval" on page 222)
1)
The Technical Handbook and Instructions for Use/data sheets for the sensors used and the PC
software CC-Vision for Dräger X-am 5600 are available on CD.
See also attached Instructions for Use and Data Sheets of the sensors used.
The Instructions for Use/data sheets for the sensors used can also be downloaded at the
following Internet address:
http://www.draeger.com/DE/de/products/gas_detection/portable/sensors/
cin_draegersensor_xxs.jsp
1)
.
o
o
C (–20 to +40
C for NiMH single cells

Publicité

Table des Matières
loading

Table des Matières