Replacing the batteries / rechargeable batteries
Danger of explosion!
Do not throw used batteries into fire or try to open them by force.
Dispose of the batteries in accordance with national regulations.
Danger of explosion!
Do not replace the batteries / rechargeable batteries in areas where there is a
danger of explosion.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
Alkaline batteries – T4 – (not rechargeable!)
Energizer No. E91, Energizer No. EN91 (Industrial),
Varta Type 4106 (power one) or
Varta Type 4006 (industrial)
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC (1800) max. 40
Switching off the instrument:
Press and hold the [ OK ] key and the [ + ] key simultaneously.
Loosen the screw on the power pack and remove the power pack.
For the battery holder (order no. 83 18 703):
Replace alkaline batteries or NiMH rechargeable batteries. Ensure correct
polarity.
For the NiMH power pack T4 (order no. 83 18 704):
Completely replace the power pack.
Insert the power pack into the instrument and tighten the screw, the
instrument switches on automatically.
WARNING
WARNING
o
C ambient temperature.
Charge the instrument with NiMH power pack T4
(83 18 704)
Danger of explosion!
Do not charge underground or in explosion-hazard areas!
The chargers are not designed in accordance with the regulations for firedamp
and explosion protection.
Even if the instrument is not used, we recommend that you store the instrument
in the charger!
Insert the switched off instrument into the charger.
Display LED on the charger:
To maintain the lifetime of the batteries, charging is temperature controlled and
only performed in a temperature range of 5 to 35
range is left, the charging is automatically interrupted and automatically
continued after the temperature range has been reached again. The charging
time is typically 4 hours. A new NiMH power pack reaches its full capacity after
three complete load/unload cycles. Never store the instrument for extended
periods (maximum 2 months) without being connected to a power source, as
this drains the internal buffer battery.