Do not disassemble the spring unit as it is under pressure. There is a risk of injury when opening the
spring unit sleeve. The spring unit and the O-ring for the NEURO SWING H
not be greased.
8.4
Securing the Screws
The screws are secured after the orthosis has been produced and tried on and before it is handed over to the
patient.
1 Secure the screw for the joint case (fig. 6) with the torque corresponding to the system width and
LOCTITE® 243 medium strength.
2 Let the adhesive harden (final strength after approx. 24 hours).
Screw for Joint Case
S1 (screw 1, axle screw)
The screw of the joint case is not secured with the necessary torque at delivery. You can also find infor-
mation on the torque on the cover disc of the system joint.
9.
Adjustment Options on the Orthosis
An orthosis with adjustable system ankle joints (fig. 8) can be individually
adapted to the pathological gait. The adjustments described in paragraph
9.1 do not influence each other and can be changed separately.
Mind the correct adjustment of the dorsiflexion stop when mount-
ing the system ankle joint. It is decisive for the entire alignment of
the orthosis.
9.1
Adjustments on the Spring Unit
There are spring units with disc springs (2) and pressure springs (3). The alignment of the orthosis can be
adjusted by screwing and unscrewing the spring units (1; fig. 8). The spring force can be changed with spring
units in different strengths.
System Width
12mm
14mm
3Nm
4Nm
O system ankle joint must
2
16mm
20mm
4Nm
4Nm
1
2
1
3
fig. 8
23