8. Input the passcode 0000
using the number buttons
to confirm that a new
NO scrubber is installed
9. Select OK

Disposal of used/expired products

• Used or expired NIOX MINO
recycled according to the local recycling program
for electronic equipment. Note that there is a
LiMnO
battery inside the Sensor unit*.
2
• An expired NIOX MINO should be recycled,
according to the local program for electronic
equipment. Note that there is a LiMnO
battery inside the instrument*.
• The NO scrubber contains potassium
permanganate and should be disposed of as
hazardous waste in accordance with the local
waste disposal regulations.
NIOX MINO is RoHS compliant.
* The batteries are not user-replaceable parts.

Return shipments

Contact your local Aerocrine representative or
Aerocrine AB.

Support

Please contact your local Aerocrine representative
or Aerocrine Customer Support if you encounter
problems, which you can not solve with the actions
stated in this manual.
For contact details, see back-page, and provide the
following information:
• Your name, address and telephone number.
• Serial No. (both instrument and Sensor).
• Problem description (as thorough as possible).
• Alert codes or lists.

Limited warranty

Aerocrine AB provides a Limited Warranty for this
instrument and original accessories delivered with
this instrument. Conditions are defined when the
items are purchased.
Do NOT try to repair the instrument. It is NOT
permitted to open the instrument. Any attempt
will make the warranty invalid and performance
according to the specifications can not be
guaranteed.
12
Sensors should be
®
backup
2

Troubleshooting

Warning
Action
The inhalation was too weak
to initiate a measurement or an
exhalation into the instrument
was performed prior to an
inhalation - Stop the procedure
immediately when this warning
appears. Wait until the main
menu screen is displayed and
repeat the inhalation with a
stronger inhalation force.
NO scrubber reminder.
The symbol is shown at first
start-up of the instrument
as a reminder to insert and
set the software for a new
NO scrubber. See Installation
and set-up section page 4.
NO scrubber almost expired -
Order a new NO scrubber.
The symbol is shown when
10% of the measurements are
remaining or 2 weeks before
expiration date and continue
until the NO scrubber has
expired. A NO scrubber can be
used for 1000 measurements
or 1 year. Refer to the Change
NO scrubber section page 11.
No Sensor connected -
Connect a Sensor.
Sensor almost expired – Order
a new Sensor.
The symbol is shown when
10% of the measurements
are left or 2 weeks before
expiration date and continue
until the Sensor has expired.
Refer to the Change Sensor
section page 11.