Section 6 Startup
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Personal injury hazard. Never use the instrument without the syringe cover installed.
6.1 Configure the instrument
1. From the main menu, select Settings.
2. Select an option, then push Select.
Option
Applications
Operators
Date + Time
Brightness
Sounds
Language
Network
Info
Restore Defaults Sets the instrument to the default configuration.
Options
Security
3. Push Back.
6.1.1 Cell standby
Make sure that the measurement cell is prepared before selecting an application, because the cell
standby starts automatically.
Cell standby dries the measurement cell automatically using a regulated addition of the titrant as
defined in the application. When the cell standby option is active:
• Cell standby starts automatically as soon as an application is selected from the home screen
• The stirring speed is the speed specified for the application
C A U T I O N
C A U T I O N
C A U T I O N
Description
Changes, makes copies, exports and removes application data. Make sure
that the duplication function does not make more than five applications.
Adds, changes and removes operators.
Sets the instrument date and time.
Sets the brightness of the display.
Sets the sound options.
Sets the language.
Give a name to the instrument. If a printer is connected, this name is printed
on the hard-copy output. Restart the instrument if the name is changed.
Shows information about the instrument and the attached hardware.
Sets the application parameters view to expert mode. When the instrument
is set to off, sets the syringe to empty into the titrant bottle. Changes the
temperature display unit. Set the cell standby parameter (refer to
standby
on page 19). Prints the measurement and derivative curves if a
printer is connected. Specify if a balance is connected.
Change the password and specify which options are password protected.
Cell
English 19