Hach PAT700 Manuel D'utilisation De Base page 23

Masquer les pouces Voir aussi pour PAT700:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
r
= average TOC response for the benzoquinone standard
ss
8. When the suitability test is completed, the results are shown. If a printer is attached, the report
prints automatically. Select the necessary action to complete the validation.
• Push CHECK to accept the system suitability test and to remove the bottles. This option will
not show if the Autoaccept setting is enabled.
• Push CANCEL to repeat the system suitability test and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Measure a grab sample
Use the grab sample test to measure samples taken from other locations. Grab samples do not need
bottles with RFID tags.
1. Push << to show the sliding menu.
2. Select BOTTLE MODE, then GRAB SAMPLE.
3. Obey the screen prompts to enter the values.
Note: To save the new settings as the default, push SAVE AS DEFAULT.
4. Install the standards in the order shown. If the bottle bay door closes too soon, push UNLOCK to
open the door.
5. Set the volume of each sample bottle.
6. Set an ID for each bottle.
7. Push START.
Note: When the measurement is complete, the results show on the screen. If a printer is attached, the report
prints automatically.
8. Select the necessary action to complete the calibration.
• Push CHECK to accept the measurement and remove the bottles. This option does not show if
the autoaccept setting is enabled.
• Push CANCEL to repeat the measurement and discard the results.
• Push EXPORT to send the data to the supplied USB flash drive.
Manage measurement data
Measurement data can be reviewed on the instrument, printed, exported and filtered.
1. Push << to show the sliding menu.
2. Select DATA REVIEW. The data review window shows. Icon command options show on the
bottom of the screen.
Set the data filter
Data can be filtered by test type, date range or special ranges.
1. Push << to open the sliding menu.
2. Select DATA REVIEW, then FILTER.
3. Select all of the data types that apply. All types are selected by default.
4. Select time range or special range. Special range options sets the data to sort first by type, then
by time. The time range option sets the data to sort based on the specified date range.
Option
Last accepted
Last done
Last five
Description
Includes only the last accepted tests for the selected data types
Includes only the last completed tests for the selected data types
Includes only the last five completed tests for the selected data types
English 23

Publicité

Table des Matières
loading

Table des Matières