Hach PAT700 Manuel D'utilisation De Base page 28

Masquer les pouces Voir aussi pour PAT700:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
Operation may continue when a warning condition is present. Refer to
contact technical support.
Code
Message
SC100–SC103
IncorrectBottleLoadedinPositionX Examine and change the bottle as necessary.
SC104–SC107
NotEnoughFluidinBottleX
SC127
UvLampLifeExceeded
SC153–SC156
BottleXMarkedAsUsed
SC165
LampSwitch
SC172
RougeDetected
SC178
ExcursionCaptured
SC181
BottleTestCancelled
SC182–SC185
BottleXExpired
Measurement alarm codes
Measurement alarms occur when TOC or conductivity levels in the process water are more than the
set limits. Refer to
Table
The digital outputs that identify a TOC alarm or conductivity alarm change their state when a
measurement alarm is present. The three analog outputs continue to represent the last measured
value when a measurement alarm is present.
Code
Message
SC161
TOCAlarm
SC162
ConductivityAlarm
28 English
Table 12 Warning alarm codes
13.
Table 13 Measurement alarm conditions
Description
The measured TOC is more than the user-defined TOC level.
The measured conductivity is more than the USP645 conductivity level.
Resolution
Adjust the fluid level before the test.
Replace the lamp.
Replace the bottle with the correct standard.
One lamp is weak or has failed. Replace the lamp.
One or more of the 500 ppb standards took longer than
expected to oxidize. Clean the cell.
Remove the sample from the analyzer.
Start the test again.
Replace the bottle with the correct standard.
Table
12. For codes not listed,

Publicité

Table des Matières
loading

Table des Matières