- Wait for the procedure to finish.
Warning!
All the data acquisition operations take
place automatically if the transducer values
fall within the established tolerance range;
if this is not the case, the procedure halts
and signals the out-of-range condition.
Electronic Calibration Check
- This window is used to check the meas-
uring sensors calibration (or the value
read from several transducers including
possible electronic corrections calculated
and saved during the zero and full scale
calibration).
- The window displays the calibrator with
the measuring sensors fitted, each in its
own position.
- The data for the current check are dis-
played alongside each measuring sensor.
- The values outside the tolerance thresh-
olds are displayed in red, whereas the
values within the tolerance thresholds
are displayed in green.
The following keys are active in this window:
F1 displays the on-line "Help" function
F2
goes back to the previous
screen
F11
displays the menu with the
relative functions.
These are the menu options:
• Start context (goes back to the starting
screen)
• Print (prints the values found in the cali-
bration check)
• Down (moves on to the next step in
the calibration check - for step-by-step
performance of the zero and full scale
calibration)
104
• Up (moves back one step in the calibra-
tion check)
• Help (recalls the Help screen relative to
the current screen)
Warning!
If some values are outside the permitted
tolerance ranges the zero and full scale cali-
bration procedure should be performed.
Before performing the zero- and full scale
setting procedure, it is always important
to check that the mechanical calibration
is correct.
Mechanical Calibration Check
This window is used to check the mechani-
cal calibration of the measuring sensors (or
the value read from several transducers
without the influence of possible electronic
corrections).
The window shows the calibrator with the
measuring sensors fitted, each in its own
position.
The data for the current check are displayed
alongside each measuring sensor.
The values outside the tolerance thresh-
olds are displayed in red, whereas the
values within the tolerance thresholds are
displayed in green.
The following keys are active in this window:
F1
displays the on-line "Help"
function
F2
goes back to the previous
screen
F11
displays the menu with the
relative functions.
These are the menu options:
• Start context (goes back to the starting
screen)
• Print (prints the values found in the
calibration check)
• Down (moves on to the next step in
User Manual WA 780 - Cura S 780