If the equipment fails to operate correctly, the table below
suggests some simple checks and corrective actions. If these
steps fail to resolve the problem, contact Arjo or an approved
service agent.
Symptom
"Beeping" sound when
using the bed
One or more bed
functions inoperative
Bed is difficult to
manoeuvre
All indicators on ACP lit
or flashing
Mattress platform
cannot be lowered
Error code E300 is
displayed
Mattress platform
cannot be lowered and
error code AES is
displayed
Error code E410
62
Possible Cause
Bed is operating from
the backup battery
Function(s) locked on
ACP
Brake pedals in "steer"
position
Duty cycle of electrical
system exceeded
Height control software
error
Control button pressed
for more than 90
seconds
Anti-Entrapment system
activated
Service Error
Action
Check the power supply
cord is plugged in and
the electricity supply is
OK
Check fuse in mains
plug (where fitted)
Unlock function(s) on
ACP
Place brake pedals in
the "free" position
Refer to the section
"Duty cycle lockout" on
page 42
Raise the mattress
platform to maximum
height to reset software
Remove pressure from
control buttons. If error
code does not clear call
an Arjo approved
service engineer
Remove obstruction(s)
from below the mattress
platform. If bed still
cannot be lowered, call
an Arjo approved
service engineer
Call an Arjo approved
service engineer