Monitor Bed Status Through The Rs232 Connection - arjo Enterprise 9600X Notice D'utilisation

Masquer les pouces Voir aussi pour Enterprise 9600X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Monitor bed status through the RS232 connection

The bed continuously transmits data about its status through
an RS232 connection at the head end of the bed (see
page 15).
The transmitted data can be recorded by a compatible device
for monitoring purposes.
NOTE
An RS232 serial cable is not supplied by Arjo.
Data is transmitted every ten seconds and includes the
following information:
This information can be stored in a database.
WARNING
Data retrieved from the RS232 connection is not intended to be used
to make clinical decisions. All patient diagnosis, treatment and care
should be performed under the supervision of an appropriate
healthcare professional.
• Patient weight
• Bed exit alarm (on or off)
• Backrest Angle in degrees relative to horizontal
• Bed height at low position (yes/no)
• Side rail position (up/down)
• Brake position (on/off)