Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601-1-2 test
Immunity test
level
Electrostatic
±2kV, ±4kV, ±8kV,
discharge (ESD)
±15kV air
EN 61000-4-2
±8kV contact
3V in 0,15 MHz
Conducted
to 80 MHz
disturbances
6V in ISM and
inducted by
amateur radio bands
RF fields
between 0,15 MHz
EN 61000-4-6
and 80 MHz
80% AM at 1 kHz
Professional
Radiated RF
Healthcare
electromagnetic
environment
field
3 V/m
EN 61000-4-3
80 MHz to 2,7 GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
Proximity fields
710, 745, 780 MHz -
from RF
9V/m
wireless
810, 870, 930 MHz -
communications
28 V/m
equipment
1720, 1845, 1970,
EN 61000-4-3
2450 MHz – 28 V/m
5240,5500, 5785 MHz
- 9V/m
±1kV SIP/SOP ports
Electrical fast
±2kV AC port
transient/burst
100kHz repetition
EN 61000-4-4
frequency
Power frequency
30A/m
Magnetic field
50 Hz or 60 Hz
EN 61000-4-8
74
Compliance level
±2kV, ±4kV, ±8kV,
±15kV air
±8kV contact
3V in 0,15 MHz
to 80 MHz
6V in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80% AM at 1 kHz
Professional Healthcare
environment
3 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz -
9V/m
810, 870, 930 MHz -
28 V/m
1720, 1845, 1970,
2450 MHz - 28 V/m
5240,5500, 5785 MHz -
9V/m
±1kV SIP/SOP ports
±2kV AC port
100kHz repetition
frequency
30A/m
50 Hz
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material the relative
humidity level should be at
least 30%
Portable and mobile RF
communications equipment
should be used no closer to
any part of the product,
including cables, than 1.0m, if
the transmitter's output power
a
rating exceeds 1W
.
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey, should be less than
the compliance level in each
b
frequency range
.
Interference may occur in the
vicinity of equipment marked
with this symbol:
Mains power supply should
be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.