remaining soil paying particular attention to the inside
of the forceps hinge, serration grips and pin. Thoroughly
rinse in fresh running water and/or distilled water to
remove all traces of detergents and soil. Forceps should
be inspected for cleanliness and damage, ensure they
function correctly and are clean prior to sterilization.
Sterilization
The tissue everting forceps may be steam sterilized
in a gravity displacement or prevacuum autoclave).
] It is recommended that the tissue everting forceps
be sterilized in accordance with one of the following
sterilization times and temperatures:
Gravity-Displacement Steam Sterilization
Wrapped Forceps
Temperature: 270-275°F (132-135°C)
Exposure Time: 10 minutes
Gravity-Displacement Flash Cycle
Unwrapped Forceps
Temperature 270-275°F (132-135°C)
Exposure Time: 10 minutes
Pre-Vacuum Steam Sterilization
Wrapped Forceps
Temperature: 270-275°F (132-135°C)
Exposure Time: 4 minutes
Pre-Vacuum Flash Cycle
Unwrapped Forceps
Temperature 270-275°F (132-135°C)
Exposure Time: 4 minutes
Notices: Limited Product Warranty; Limitation
Of Remedies
LeMaitre Vascular, Inc. warrants that reasonable care
has been used in the manufacture of this device. Except
as explicitly provided herein, LEMAITRE VASCULAR (AS
USED IN THIS SECTION, SUCH TERM INCLUDES LEMAITRE
VASCULAR, INC., ITS AFFILIATES, AND THEIR RESPECTIVE
EMPLOYEES, OFFICERS, DIRECTORS, MANAGERS, AND
AGENTS) MAKES NO EXPRESS OR IMPLIED WARRANTIES
WITH RESPECT TO THIS DEVICE, WHETHER ARISING
BY OPERATION OF LAW OR OTHERWISE (INCLUDING,
WITHOUT LIMITATION, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE) AND HEREBY DISCLAIMS THE SAME. LeMaitre
Vascular makes no representation regarding the
suitability for any particular treatment in which
this device is used, which determination is the sole
responsibility of the purchaser. This limited warranty
does not apply to the extent of any abuse or misuse of,
or failure to properly store, this device by the purchaser
or any third party. The sole remedy for a breach of this
limited warranty shall be replacement of, or refund
of the purchase price for, this device (at LeMaitre
Vascular's sole option) following the purchaser's return
of the device to LeMaitre Vascular. This warranty shall