Troubleshooting; Replacement Parts And Accessories - Hach Flo-Station Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Troubleshooting

When a problem occurs, identify if the problem is with the instrument, cables or sensor.
1. Make sure that power is applied to the instrument.
2. Make sure that all the cable connections are secure.
3. Remove debris from the sensor.
4. Identify if the problem is with one measurement channel or if the sensor does not transmit.
5. If the sensor does not transmit:
a. Look for moisture on the sensor connector.
b. Dry the sensor connector as necessary.
c. Make sure that the sensor connector is tightened securely.
6. Record all error codes that occur on the instrument, then contact the manufacturer.
7. If the instrument stops operation as a result of a power fluctuation or a reason not known, push
the red power reset button to start the microprocessor again. Refer to
Figure 6
on page 12.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Cable, RS232, 3 m (12 ft)
Filter, hydrophobic
Fuse, 5 x 20 mm, 1 A, 250 VAC, antisurge
Power cable, 115 VAC, 10 A, 2.4 m (8 ft)
Accessories
Description
Desiccant cartridge, moist environment
Desiccant cartridge, moist environment, low maintenance
Figure 5
on page 11 or
Item no.
570007301
151000201
015804
24500M
Item no.
55032
5027