Troubleshooting

Error
Sipper module not
There is no USB
connected. Please
connection between the
check cable.
sipper and the
turbidimeter.
Please check sipper
The pump tubing is not
and tube.
correctly installed.

Replacement parts and accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Cable, USB type AB, 1 m (3.3 ft)
Cleaning lid
Cleaning lid, screws (3x)
Flow-thru head, TU5200
Seal, cleaning lid
Seal, process vial
SIP 10 sipper unit with LZV940
Tubing, pump (Lagoprene) pre-assembled with peristaltic pump cover and
connections
Tubing kit, includes Tygon tubing 1.70 m (5.6 ft), pump tubing (Lagoprene) pre-
assembled with peristaltic pump cover and connections
Vial, process
Vial nut
Vial replacement tool
Accessories
Description
Micro fiber cloth (for vial cleaning)
Vial wiper
16 English
Description
Examine the USB cable. Make sure that the cable length is
not more than 1 m (3.3 ft). Make sure that the USB cable is
connected to the turbidimeter and the sipper.
Loosen the peristaltic pump cover, then install it again. Refer
to the illustrated steps in
on page 14.
Operate the pump for 15 seconds. Push down the lever and
make sure that the tube is correctly installed around the
rollers.
W A R N I N G
Solution
Replace the pump tubing
Item no.
LZQ104
LZY919
LZY921
LZV967
LZY914
LZY918
LQV157.99.40001
LZV877
LZV940
LZY834
LZY917
LZY906
Item no.
LZY945
LZY903