Hach Flo-Station Manuel D'utilisation page 11

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Figure 5 Wiring for AC power
1 Fuse holder
2 Power reset button
3 Ground wire
4 Neutral wire
Connection
Hot (L)
Neutral (N)
Protective earth ground (G)
DC power
Electrocution hazard. Do not connect AC power directly to a DC powered instrument.
1. Remove the bottom cover. Refer to
2. Remove approximately 6 mm (¼ in.) of insulation from each wire.
3. Attach the ground wire or cable braid to the protective earth grounding block. Make sure that the
ground wire is held securely. Refer to
4. Put the negative wire and the positive wire in the DC power terminal. Refer to
5. Tighten each wire connection so that the wires are held securely.
6. Tighten the cable strain relief or conduit hub so that the cable or conduit is held securely.
7. Install the bottom cover so that the enclosure is sealed.
5 Line wire
6 AC power terminal block
7 Ferrule connector
8 Protective earth terminal
Table 1 AC wiring information
Color—North America
Black
White
Green
D A N G E R
Figure 1
on page 6.
Figure
6.
Color—EU
Brown
Blue
Green with yellow stripe
Table
2.
English 11

Publicité

Chapitres

Table des Matières
loading

Table des Matières