IV. 6. Technical characteristics
Dimensions: D x W x H: 368 x 419 x 699 mm.
Weight: 25.0 kg.
Noise level: ± 52 dBA
Flow values:
0 - 8 l/min.
Average oxygen content:
• at 2 l/min. 92 %.
• at 7 l/min. 91 %.
• at 8 l/min. 90%.
(values at 21 °C and at an atmospheric pressure of 1013 mbar). Max. recommended flow:
8 l/min. The maximum outlet pressure is 138 kPa. It is not recommended to use the NewLife
Intensity at flow values less than
2 l/min. unless utilized with the Pediatric Option Procedure.
Electrical power supply:
• 230 V - 50 Hz Europe / 115 V – 60 Hz / 230 V – 60 Hz (other countries,
depending on version)
• Use the cable provided, and check that the electrical characteristics of the mains socket used
match those indicated on the manufacturer's plate on the rear panel of the machine.
• Mean power rating: 410 watts
Class II
Type BF
2.0 amps (230 V).
4.0 amps (120 V).
Filters:
At the rear of the device: a dust filter.
Before the oxygen outlet: a product filter < 2.0 µm.

Air circulation:

A fan cools the compressor compartment.
Environmental limit conditions:
The performances of the device (especially the oxygen concentration) are quoted at 21 °C and
1013 mbar. They may change with temperature and altitude.
- The device should preferably be stored and transported in the vertical position.
- The device must only be used in the vertical position.
- Ambient temperature of between 5°C and 40 °C (operation).
- Storage temperature range from -20 °C to 60 °C.
- Relative humidity of between 0% and 95% (operation and storage).
- IPX1: Protected against dripping water (complies with the EN 60601-1 standard)