Technical specifications
Operating pressure range
Maximum single fault steady state pressure
Pressure measurement tolerance
Flow measurement tolerance
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
Sound power level
Nominal dimensions (L x W x H)
Weight
90W power supply unit
30W power supply unit
Operating temperature
Operating humidity
Operating altitude
Storage and transport temperature
Storage and transport humidity
Housing construction
Supplemental oxygen
Hypoallergenic air filter
Standard air filter
SlimLine
air tubing
TM
Standard air tubing
Air outlet
Electromagnetic compatibility
4 to 20 cm H
O
2
30 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±6 L/min or 10% of reading, whichever is greater
24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according to ISO 17510-
1:2007
34 dBA with uncertainty of 2 dBA as measured according to ISO 17510-
1:2007
153 mm x 140 mm x 86 mm
810 g
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 70W (80VA)
Maximum power consumption 110W (120VA)
Input range 100–240V, 50–60Hz,
Nominal for aircraft use 110V, 400Hz
Typical power consumption 20W (30VA)
Maximum power consumption 36W (75VA)
º
º
+5
C to +35
C
10–95% non-condensing
Sea level to 2,591 m
º
º
-20
C to +60
C
10–95% non-condensing
Flame retardant engineering thermoplastic
Recommended maximum supplemental oxygen flow: 4 L/min
Non-woven acrylic and polypropylene fibers with polypropylene carrier
Polyester fiber
Flexible plastic, 1.83 m, 15 mm inner diameter
Flexible plastic, 2 m, 19 mm inner diameter
The 22 mm conical air outlet complies with ISO 5356-1
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments. Information regarding the
electromagnetic emissions and immunity of these ResMed devices can be
found on www.resmed.com, on the Products page under Service and
Support. Click on the PDF file for your language.
5