Télécharger Imprimer la page

ResMed S9 AutoSet Mode D'emploi page 5

Masquer les pouces Voir aussi pour S9 AutoSet:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
Problem/Possible cause
Displays error message: High temperature fault, refer to user manual
Device has been left in a hot environment.
The device or air inlet is being covered by
bedding.
Air filter is blocked.
Air tubing is blocked.
Humidifier setting is too high, resulting in
accumulation of water in the air tubing.
Displays error message: Check ResMed Power Supply Unit and fully insert the connector
The DC plug is partially inserted into the back of
the device or inserted too slowly.
A non-ResMed power supply unit is connected to
the device.
The power supply unit is being covered by
bedding.
The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card
error, please remove SD card and contact service provider
SD card is not inserted correctly.
You may have removed the SD card before
settings were copied to the CPAP device.
The following message is NOT displayed on the LCD after you try to update the settings using the SD card: Settings
updated successfully, press any key
The settings were not updated.
SD card
An SD card has been supplied to gather therapy data from your S9 device and provide settings updates from your
clinician. When instructed to do so, remove your SD card, insert it in the protective folder and send it to your clinician.
For more information on removing and inserting your card refer to the S9 SD Card Protective Folder provided with your
device. Please retain the S9 SD Card Protective Folder for future use.
4
Solution
Allow to cool before re-use. Disconnect the power cord and then
reconnect it to restart the device.
Make sure that the device is free from bedding, clothes or other objects
that could cover the device or block the air inlet. Disconnect the power
cord and then reconnect it to restart the device.
Replace your air filter. Disconnect the power cord and then reconnect it to
restart the device.
Check your air tubing and remove any blockages. Disconnect the power
cord and then reconnect it to restart the device.
Turn the humidifier setting down and empty the water from the air tubing.
Fully insert the DC plug.
Remove the power supply unit and replace with a ResMed power supply
unit.
Make sure that the power supply unit is free from bedding, clothes or
other objects that could cover it.
Ensure that the SD card is inserted correctly.
Reinsert the SD card and wait for the home screen or the Settings
updated successfully, press any key message to appear on
the LCD.
Note: This message only appears once. If you re-insert the SD card after
you have updated your settings, the message will not be re-displayed.
Contact your clinician/service provider immediately.

Publicité

loading

Ce manuel est également adapté pour:

S9 elite