System Specifications - Sartorius Stedim Biotech arium 61316 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

System Specifications

Unit Measurements:
Note
38 cm (15") in front of the unit must be available to open the doors and 10 cm (4") on each side for connections.
System Weight:
Operating Weight:
Inlet Water Requirement:
Technical Data:
Tolerance limit: product water
after RO modules
Programmable times printer in
the print mode
Rejection rate
Display conductivity|resistivity
value of product water
Operating Specifications:
Flow Rate Product|Permeate:
RO Reject|Concentrate:
Recovery RO, approximately:
Note
Permeate flow will change with water temperature. Lower temperature = less permeate.
Electrical:
Voltage
Frequency
Power Consumption
Fuses
Ambient Operating Temperature:
Ambient Storage:
43 cm (17") width, 48 cm (19") height and 33.4 cm (13.5") depth
14 kg (31 lb)
20 kg (44 lb)
Only use potable water as feed water according to national laws and regulations
(Max. total hardness: 360 ppm as CaCO
In the case of hard water, it makes sense to soften the feed water
Iron maximum: 0.1 ppm
Colloid index or SDI: <3
Temperature range: 2°C - 35°C
Pressure: min. 1 bar (15 psi), max 6.8 bar (100 psi)
Adjustable from 10–80 μS/cm or 0.013–0.1 MO + cm
From 1...59 min
From 85 to 99%
μS/cm or corresponding MO + cm
compensated to 25°C
Monovalent ion rejection rate: up to 98%
Polyvalent ions rejection rate: up to 99%
Microorganism rejection rate: >99%
Particle rejection rate: >99%
TOC: <100 ppb or >99% for MW >300
up to 16 l/h @ 25°C
up to 12 l/h
55% (Depends on feed water quality and temperature)
100–240 Vac
47–63 Hz single phase
74 VA max
2 (neutral conductor|phase), 4 AT, 5+20 mm
5–30°C, 80% relative humidity, non-condensing
Temperature 5–45°C, 80% relative humidity, non-condensing
, max conductivity 1,500 μS/cm)