Tank Cleaning
Note!
The arium
®
tank should be cleaned
prior to new installation and at
regular intervals.
To remind you to perform cleaning,
the message Timeout "Tank Saniti-
zation" will appear on the display
according to the factory setting.
This message will blink on your
display until a complete cleaning
cycle has been performed.
The audio alarm can be stopped by
switching the Enter button.
At the end of the sanitization cycle,
this reminder function will reset
automatically.
!
Warning!
The tank system must be cleaned
with a special cleaning|sanitization
solution.
Wear protective eyeglasses and rub-
ber|latex gloves when performing
the sanitization procedure. Labora-
tory chemical safety procedures
should be followed when handling
any laboratory chemicals.
Have the following ready prior to
cleaning:
– Cleaning|Sanitization kit with
Order no. 611CDS
To clean the tank:
1. Press Menu to activate the Main
Menu.
MENU
Maintenance
Setup
Info
2. Select <Maintenance> and press
Enter to confirm. The display will
show the following:
Maintenance
Repl. RO modules
Repl. Prefilter
Sanitization
22
3. If reguire, use the arrow key to
scroll down and select <Sanitiza-
tion> in the menu. Press Enter to
confirm.
Sanitization
System
Tank
4. Select <Tank> and press Enter to
confirm. The display will show the
following:
Sanitization
Tank
Continue?
Yes
No
Note!
The program can no longer be
interrupted, once the safety prompt
has been confirmed.
5. Choose <Yes> to proceed tank
cleaning and press Enter to con-
firm. Follow the instructions that
appear in the display and push
Enter after each step is completed.
Note!
For the injection of the sanitization
solution unscrew the luer end cap
from the connection of the adapter.
Inject the cleaning|sanitization
liquid through the port into the
system (see picture) and then
remove the syringe. Afterwards
reatach the luer end cap.
Note!
After sanitization check to make
sure that all tubes are securely
installed and leaktight.
After all work is completed,
the unit is operational again.