Initial Operation

Before the initial operation of the
arium
®
61316, the previously
installed consumables must be put
through a 60-minute rinsing
process.
The arium system recognizes
automatically whether or not it
is an initial operation when the
device is switched on.
The rinsing process is automatically
activated when the device is switched
on ("I") in the case of initial opera-
tion.
Please follow the notes that appear
on the display.
Note!
All installation steps have to be
completed before you switch on
the system power and pump in
feed water.
1. The main power switch on the
left-hand side of the housing
must be on ("I").
The arium
®
system will then
perform a system check.
The arium
®
system display will show
the following information (in Eng-
lish upon initial operation):
SYSTEM ARIUM RO
S/N xxxxxxxxxx
System Check OK
V1.XX
The display will show the model
type of the system, the serial
number and the software version.
Following this, there is a further
system test for initial operation.
In the case of initial operation, the
system automatically changes to
the rinsing process.
Note!
If initial operation has already been
carried out, the system misses out
the rinsing process.
Following this, the system carries
out internal initialization as well
as an inspection of the inlet pres-
sure (minimum 1 bar, maximum
6.8 bar).
Initialization
____ Tank level
Time
Date
When there is a tank pressure of
< 2.5 bar (in the case of depressur-
ized tank volume < 50%), the sys-
tem then changes to the "operate"
mode or, in the case of a tank
pressure > 2.5 bar (in the case of
depressurized tank volume > 50%),
to the "ready" mode.
The following display appears in the
"operate" mode:
XX
%
Rejection Rate
XX μS/cm comp.
_ _ _ _ _ _ _ _ _ _ Tank Level
Time
Basic settings
After the arium
®
61316 rinsing
process, set the language for the
display text, date and time, the
desired measurement units, the
limit value and the tank system.
Note!
In some menus, not all menu items
are visible on the display. Repeated-
ly press the arrow key
to dis-
play the required language.
Display Language
This option allows you to set the
language for the menu texts in the
display. The system is factory set to
English.
Proceed as follows:
1. Press Menu:
The display will show the following
information:
MENU
Date
2. 2. Select <Setup> and press Enter.
The following display will appear:
Setup
3. Use the arrow key to scroll down,
select <Language> and press Enter.
The following display will appear:
Language
Use the arrow keys to select the
desired language and press Enter to
confirm.
The display will then switch to
"operate" mode.
Maintenance
Setup
Info
Time/Date
Print Mode
Set Point
Meas.Unit
Tank
Language
Audio Alarm
English
German
17