CS-3000 RF Generator Unit Operator Manual
Mode Button (4)
The Mode button is used to adjust the operating mode between Standby, Power Control,
and Diagnostic Evaluation. Pressing the Mode button during the application of RF energy
will be ignored by the system.
RF ON/OFF Button (1)
The RF ON/OFF button works in parallel with the Footswitch (see Footswitch description).
During Power Control Mode, the RF ON/OFF button controls the operation of the RF
generator by initiating or terminating RF energy.
RF ON/OFF Indicator (8)
The RF ON/OFF indicator lights up when RF energy is being transmitted. RF energy
transmission includes the periodic delivery of energy to measure impedance, even when the
RF generator has not been activated. However, when RF energy is being transmitted at the
power capable of causing coagulation, the RF ON/OFF indicator light is constantly
illuminated.
Connections
Front Panel Connectors
The Footswitch, RF Coagulation Cable for the Device, Indifferent, Dispersive Electrode and
Diagnostic Port for Sensing Cable CSK-2010 connectors allow interfacing with accessory
devices.
Footswitch (11)
Depress the footswitch to begin delivering RF energy. To terminate the energy delivery and
reset the Time during operation, depress the footswitch again.
Indifferent, Dispersive Electrode (10)
The indifferent, dispersive electrode provides a path for the electrical current through the
patient and back to the generator. It is important to properly attach the indifferent, dispersive
electrode to the patient per manufacturer's instructions (see "Setup and Operation"). The
indifferent, dispersive electrode is for single use only.
Sensing Cable Connection (12)
The Sensing Cable Interface (Generator 'diagnostic' port) allows the connection of the
sensing electrodes from the EPi-Sense device to the external EP Sensing (EKG)
equipment, with the use of the CSK-2010 cable.
RF Coagulation Cable for the Device (9)
AtriCure manufactures the coagulation devices and RF coagulation cable for exclusive use
with the Model CS-3000 RF Generator Unit. Refer to the two Figures (5 & 6) below, for the
appropriate setup.
CSK Type Devices, CSK-2000 Cable and RF Generator Setup
Refer to the drawing below for the appropriate setup of the CSK devices with the CSK-2000
cable and the CS-3000 generator.
Figure 5. CSK Type Devices, CSK-2000 Cable and RF Generator Setup
CDK Type Devices, CSK-2030 or CSK-2010, CSK-2000 Cable and RF
Generator Setup
Refer to the drawing below for the appropriate setup of the CDK devices with the CSK-2000
cable, CSK-2030 or CSK-2010 cable and the CS-3000 generator.
AtriCure, Inc.
Figure 6A. CDK Type Devices, CSK-2000, CSK-2030 with over molded yoke
and RF Generator Setup
Figure 6B. CDK Type Devices, CSK-2000, CSK-2030 with enclosure box and
RF Generator Setup
Note: Pin Call out for CSK-2010 and CSK-2030 are the same
D1 = Distal Electrode 1 = Shrouded Pin #1; P1 = Proximal Electrode 1 = Shrouded Pin #3;
D2 = Distal Electrode 2 = Shrouded Pin #2; P2 = Proximal Electrode 2 = Shrouded Pin #4
Figure 6C. CDK Type Devices, CSK-2000, CSK-2010 Cable and RF Generator
Setup
CAUTION: Ensure that the EP Sensing (EKG) equipment comply with IEC 60601-2-25
for protection from high frequency surgical interference to avoid disruption or incorrect
sensing.
2021/09
English
LBL-2123 Rev. G
Page 9 / 92