Guidance & Manufacturer's Declaration- Electromagnetic Immunity - nContact CS-3000 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 42
CS-3000 RF Generator Unit Operator Manual
Guidance & Manufacturer's Declaration– Electromagnetic Immunity
Product complies with the requirements of directive 93/42/EEC.
Immunity Test
± 2 kV, ± 4 kV, ± 8 kV Contact for
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV Air
± 2 kV @ 100 kHz repetition frequency
Electrical Fast Transient /
Burst
IEC 61000-4-4
± 2 kV @ 100 kHz repetition frequency
Surge
± 0,5 kV, ± 1 kV, ± 2 kV Line-to-Ground
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
Voltage Dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power Frequency
(50/60 Hz) Magnetic field
IEC 61000-4-8
Immunity test
Immunity from Radiated
RF EM Fields including
proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
AtriCure, Inc.
IEC 60601 Test Level
conductive parts
Discharge for insulated parts
for power supply lines
for input/output lines
Power inputs
± 0,5 kV, ± 1 kV Line-to-Line
Signal input/outputs:
± 2 kV Line-to-Ground
0,15 MHz – 80 MHz
3V, 80 % AM at 1 kHz
ISM bands between
0,15 MHz and 80 MHz
6V, 80 % AM at 1 kHz
Voltage Dips:
0 % UT; 0,5 cycle
and 315° phase angles
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
NOTE: UT is the ac mains voltage prior to application of the test level.
Band (MHz)
150 kHz to 80 MHz
80 MHz – 2,7 GHz
380 –390
430 – 470
704 – 787
800 – 960
1,700 – 1,990
2,400 – 2,570
5,100 – 5,800
2021/09
Compliance Level
± 8 kV CD
± 15 kV AD
± 2 kV @ 100 kHz repetition frequency
for power supply lines
± 2 kV @ 100 kHz repetition frequency
for input/output lines
Power inputs
± 0,5 kV, ± 1 kV Line-to-Line
± 0,5 kV, ± 1 kV, ± 2 kV Line-to-Ground
Signal input/outputs:
± 2 kV Line-to-Ground
0,15 MHz – 80 MHz
3V, 80 % AM at 1 kHz
ISM bands between
0,15 MHz and 80 MHz
6V, 80 % AM at 1 kHz
Voltage Dips:
0 % UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315° phase angles
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
Wireless Service
General
General
TETRA 400
GMRS 460, FRS 460
LTE Band 13, 1
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850,
LTE Band 5
GSM 1800; CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS
Bluetooth, WLAN,
802.11 b/g/n, RFID 2450,
LTE Band 7
WLAN 802.11 a/n
Electromagnetic Environment -Guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the CS-3000 RF Generator
requires continued operation during power mains interruptions, it is
recommended that the CS-3000 RF Generator be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
Immunity Test Level (V/m)
< 3
3
27
28
9
28
28
28
9
English
LBL-2123 Rev. G
Compliance Test Level
(V/m)
< 3
3
27
28
9
28
28
28
9
Page 5 / 92