CS-3000 RF Generator Unit Operator Manual
Portable and mobile RF communications equipment should be used no closer to any part of the CS-3000 RF Generator including cables, than the recommended separation distance
calculated from the equation:
d = 6/E × √P
Where:
d is the separation in meters
P is the maximum output power rating of the transmitter in watts (W) according to the Transmitter Manufacturer
E is the Compliance Test Level indicated above.
Interference may occur in the vicinity of equipment marked with the following symbol:
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the ASU System or any of its components are used exceeds the applicable RF compliance level
above, the ASU System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating components or the entire ASU System.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
AtriCure, Inc.
2021/09
English
LBL-2123 Rev. G