Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Problem/
Possible cause/remedy
question
What should
1. On infants and children, the mask
be taken into
should cover the mouth and nose
account when
to ensure effective inhalation.
using the
2. On children, the mask should also
device with
cover both the nose and mouth.
infants and
It is not a good idea to carry out
children?
nebulisation on someone who
is sleeping, as in this case not
enough of the medication will
reach the lungs.
Note: Inhalation should only be car-
ried out under the supervision of an
adult and with their assistance and
the child should not be left alone.
Inhalation us-
This is due to technical reasons. You
ing the mask
breathe less medication per breath
takes longer.
through the holes of the mask than
via the mouthpiece. The aerosol is
mixed with ambient air via the holes.
Why should
There are two reasons for this:
the atomiser
1. To ensure a therapeutically effec-
be replaced
tive particle spectrum, the nozzle
regularly?
hole must not exceed a specific
diameter. Due to mechanical and
thermal stresses, the plastic is
subject to a certain degree of
wear. The atomiser insert [11] is
particularly sensitive. This can also
alter the composition of the aero-
sol droplets, which directly affects
the effectiveness of the treatment.
2. Regularly changing the atomiser
is also recommended for hygiene
reasons.
Should each
Yes, this is absolutely essential for
person have
hygiene reasons.
their own
atomiser?

11. Technical specifications

Model
SR IH 1
Type
IH 60
Dimensions
(W x H x D)
110 x 62 x 47 mm
Weight
222 g
Operating
pressure
Approx. 0.25 to 0.5 bar
Atomiser filling
Min. 2 ml
volume
Max. 6 ml
Medication
flow rate
Approx. 0.25 ml/min
Sound pressure
Max. 45 dBA
(acc. to DIN EN 13544-1 section 26)
Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A
Output: 5 V
Expected
400 h
service life
Operating
Temperature: +10 °C to +40 °C
conditions
Relative humidity: 10% to 95%
Ambient pressure: 700 to 1060 hPa
Storage and
Temperature: -20 °C to +60 °C
transportation
Relative humidity: 10% to 95%
conditions
Ambient pressure: 700 to 1060 hPa
Aerosol properties 1) Aerosol delivery: 0.15 ml
2) Aerosol delivery rate: 0.03 ml/min
3) Particle size (MMAD): 4.12 µm
Battery:
Capacity
2570 mAh
Nominal voltage
3.7 V
Type designation
Li-Ion
The serial number is located on the device or in the bat-
tery compartment.
Subject to technical changes.
23
; 2 A

Publicité

Table des Matières
loading