Hach PHOSPHAX sigma Manuel D'utilisation De Base page 137

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 203
CURRENT 1
CURRENT 2
SETTINGS
P/OP (number of measurement cycles per parameter each 0–10)
RELAY-MIN (lower limit-value contact 0.00 to [current range P] setting)
RELAY-MAX (upper limit-value contact 0.00 to [current range P] setting)
SIGMATAX (sample feed via SIGMATAX: yes, no)
DATE (setting of the date)
TIME (setting of the time)
CLEANING (automatic cleaning off, 6 h, 12 h, 1–7D, start time for each hourly --> last Display of last cleaning)
CALIBRATION (automatic calibration off, 1–7D, start time for each hourly --> last Display of last calibration)
VERSION (program version)
PASSWORD (password protection on activation 4-digits using digits 1-4)
SERVICE STATION (place and contact partner)
BACKLIGHT (display lighting)
Current output 1
PARAMETER
(one of the four display values in the display can be assigned to current
output 1 PRD, P-oP (=PΔ), PΣ (=Total P),
oP (=Ortho-P))
CURRENT RANGE (measurement range for current output 1
0.5–5.0; 0.5–10.0–500 at [Parameter] = PRD)
CURRENT (current transfer for current output 1 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in case of fault with current output 1
off (= last measurement value), 0 mA, 20 mA)
Current output 2
PARAMETER (one of the four display values in the display
can be assigned to current output 1 PRD, P-oP (=PΔ), PΣ (=Total P),
oP (=Ortho-P))
CURRENT RANGE (measurement range for current output 1
0.5–5.0; 0.5–10.0–500 at [Parameter] = PRD)
CURRENT (current transfer for current output 2 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in case of fault with current output 2
off (= last measurement value), 0 mA, 20 mA)
English 137

Publicité

Table des Matières
loading

Table des Matières