TROUBLESHOOTING
Inflation time too long.
No cuff is connected. The cuff is too loose on the arm. The cuff is not
tight.
Place the cuff again and start a new measurement.
Incorrect cuff.
Adult cuff in neonate mode. Neonate cuff in adult mode.
Use the correct cuff.
Arrhythmia
The device determined implausible readings during the last measurement.
Take another measurement.
Artefact
The device has detected too many motion artefacts.
Take another measurement.
Excessive pressure in the cuff or measurement time too long.
Pressure was applied on the cuff during the measurement.
Do not touch the cuff and tube during the measurement.
Leak in the cuff or in the connector cable
The device recorded an unexpected decrease in pressure during the
measurement.
Use a different cuff and check the cable to the device.