Application
The activities in this chapter may only be carried out by specialist
medical personnel.
IMPORTANT: The exudate canisters of the negative pressure unit
are sterile components and can therefore be used in a sterile surgical
environment.
Inserting the exudate canister
• Carefully remove the exudate canister from the sterile peel
package.
IMPORTANT: The attached tubing piece must not be allowed to fall
onto a non-sterile surface.
• Insert the exudate canister at a slight angle into the guidance on
the negative pressure unit
• Tilt the exudate canister in the direction of the negative pressure
unit until it fully engages with the blue unlock
• Pull gently on the exudate canister to make sure that it is securely
attached to the negative pressure unit.
Removing the exudate canister
• Removing the exudate canister.
See "Connecting / removing the negative pressure unit to / from the
wound dressing set".
• Press the blue unlock key
• Slightly tilt the exudate canister and remove it.
• Properly dispose of the exudate canister. Observe local regulations.