1. Select GRAB SAMPLE>GRAB SAMPLE OUT.
2. Complete the instructions on the display.
3. Put the grab sample out tube in a clean container. The grab sample tube is located in the lower
unit of the cabinet on the left side.
4. Push down on the spigot (upper left corner of the lower compartment) for the grab sample tube to
dispense a grab sample.

Set up the system

The configuration settings can be changed in the SETUP SYSTEM menu or in the analyzer setup.
Refer to
Start the analyzer setup
1. Select SETUP SYSTEM.
2. Push menu and select SETUP SYSTEM.
3. Select an option.
Option
MEAS MODE
EDIT INTERVAL
(conditional)
MEAS UNITS
SIGNAL AVERAGE
EDIT ANALYZER
NAME
EDIT CHANNEL NAME
CONFIGURE
SEQUENCER (optional)
SAMPLE MISSING
SET DATE & TIME
DISPLAY SETUP
DISABLE REMINDERS
MANAGE DEVICES
INSTRUMENT
INFORMATION
CALCULATION
SETUP OUTPUTS
on page 6.
Description
Changes the mode of the measurement cycle. Options: interval or continuous
(default). The continuous mode measures approximately every 9 minutes.
Changes the interval time when MEAS MODE is set to interval. Options:
10–240 minutes (default = 15 minutes).
Changes the measurement units that are shown on the display and in the data
log. Options: ppb (default), ppm, mg/L, µg/L.
Selects the number of measurements used to calculate an average measurement
(1–5). This reduces variability in measurements (default = 1, no averaging).
Changes the name that is shown on the top of the measurement screen
(maximum of 16 characters).
Changes the name of the sample source that is shown on the measurement
screen (maximum of 10 characters).
Starts or stops measurements for individual sample sources. Sets the
measurement order of the sample sources when there is more than one sample
source.
Sets the operation that occurs if no sample is detected. Options: ON DELAY
(default) or OFF DELAY. ON DELAY: the instrument waits the remainder of the
measurement cycle, then goes to the next sample in the sequence. OFF DELAY:
the instrument waits 10 seconds, then goes to the next sample in the sequence.
Sets the analyzer time and date.
Changes the language. Adjusts the order measurements are shown. Adjusts the
display contrast settings.
Stops scheduled maintenance alerts for individual components. Options: tubing,
stir bar, colorimeter cell, air filter, stir motor, air relief valve, pinch valve, air
compressor, reagent valves, sample valves, standard valves, colorimeter LED, fan
filter, air check valve.
Installs or removes input modules. Refer to
more information.
Shows the analyzer information. Refer to
on page 16.
Set up variables, parameters, units and formulas for the analyzer. Refer to
the calculation
on page 10.
Selects and configures the 4–20 mA setup, the relay setup and the error hold
mode. Refer to
Configure the outputs
Manage devices
on page 10 for
View instrument information
on page 11 for more information.
Set up
English 9