Hach 5500sc Guide De L'opérateur page 13

Masquer les pouces Voir aussi pour 5500sc:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 54
3. Select ALARM in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
LOW ALARM
HIGH ALARM
LOW DEADBAND Sets the range where the relay stays on after the measured value increases above the
HIGH DEADBAND Sets the range where the relay stays on after the measured value decreases below the
OFF DELAY
ON DELAY
4. Select FEEDER CONTROL in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
PHASE
SET SETPOINT
DEADBAND
OVERFEED TIMER Sets the maximum amount of time to get to the process setpoint. When the time
OFF DELAY
ON DELAY
5. Select EVENT CONTROL in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET SETPOINT Sets the value to set the relay to on.
DEADBAND
Sets a delay so that the relay will be stable as the process value converges to the setpoint.
OnMax TIMER
Sets the maximum time the relay stays on (default = 0 min).
OffMax TIMER
Sets the maximum time the relay stays off (default = 0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the measured value (default = 0 min).
6. Select SCHEDULER in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
HOLD OUTPUTS Holds or transfers outputs for the selected channels.
Description
Sets the value to set the relay to on in response to the decreasing measured value. For
example, if the low alarm is set for 1.0 and the measured value drops to 0.9, the relay
starts.
Sets the value to set the relay to on in response to the increasing measured value. For
example, if the high alarm is set for 1.0 and the measured value increases to 1.1, the
relay starts.
low alarm value. For example, if the low alarm is set for 1.0 and the low deadband is set
for 0.5, the relay stays on between 1.0 and 1.5. The default is 5% of the range.
high alarm value. For example, if the high alarm is set for 4.0 and the high deadband is
set for 0.5, the relay stays on between 3.5 and 4.0. The default is 5% of the range.
Sets a delay time (0–300 seconds) to set the relay off (default = 5 seconds).
Sets a delay time (0–300 seconds) to set the relay on (default = 5 seconds).
Description
Specifies the relay status if the process value is larger than the setpoint. HIGH (default)
—sets the relay to on when the process value is larger than the setpoint. LOW—sets
the relay to on when the process value falls below the setpoint.
Sets the process value for the relay to toggle between the high and low value (default =
10).
Sets a delay so that the relay will be stable as the process value converges to the
setpoint.
expires and the relay does not show the setpoint, the relay is set to off. After an
overfeed alarm occurs, reset the timer manually.
Sets a delay time to set the relay to off (default = 5 seconds).
Sets a delay time to set the relay to on (default = 5 seconds).
Description
English 13

Publicité

Table des Matières
loading

Table des Matières