Maintenance Schedule - Hach CL17sc Manuel De L'utilisateur

Masquer les pouces Voir aussi pour CL17sc:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 84

7.1 Maintenance schedule

Table 6
shows the recommended schedule of maintenance tasks. Facility requirements and
operating conditions may increase the frequency of some tasks.
Task
Clean the cell
on page 20
Replace the reagent bottles
Replace the stir bar and tubing harness
Clean the screen in the Y-strainer
7.2 Clean the cell
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
Clean the cell at 1-month intervals or more frequently if necessary.
Items to collect:
Cotton swabs
1. Put on the personal protective equipment identified in the safety data sheets (MSDS/SDS).
2. Push menu, then select SENSOR SETUP > [select analyzer] > TASKS > CLEAN CELL.
Note: To stop a selected task, push home.
3. Push enter to stop measurements.
4. Select an option.
Option
HOLD
TRANSFER
5. When the status light flashes, do the illustrated steps that follow. Make sure that there is no
unwanted material in the cell. When done, push enter.
2
Clean the cell more or less frequently as necessary.
3
Refer to the instructions supplied with the Tubing Kit.
4
Refer to
Replacement parts and accessories
5
Do not use other cleaning solutions. Refer to
20 English
Table 6 Maintenance schedule
on page 21
3
on page 22
C A U T I O N
4
Description
The controller outputs are held at the last measured value.
The controller outputs change to the transfer value.
1 month
6 months
2
X
X
5.25 N Sulfuric Acid Solution
on page 27.
Replacement parts and accessories
As necessary
X
X
5
on page 27.

Publicité

Table des Matières
loading

Table des Matières