Figure 5 Connect the cable to a quick-connect fitting
3.7.2 Connect the controller to power
Connect the controller to line power by hard-wiring in conduit or wiring to a power cord. Refer to the
controller documentation for instructions.
3.7.3 Connect external devices to the controller
Connect the controller relays, analog outputs and digital inputs to external devices as necessary.
Refer to the controller documentation for instructions.
3.8 Step 7: Set up and prime the analyzer
1. Push menu, then select SENSOR SETUP > [select analyzer] > PRIME.
2. If the controller does not recognize the analyzer is connected, do the steps that follow:
a. Push menu, then select TEST/MAINT > SCAN SENSORS.
b. When the sensor is found and installed, do step
3.9 Step 8: Set the flow rate
1. Push the RIGHT arrow on the controller to show the flow rate. Sample flows through the analyzer
only when the analyzer flushes the cell. At other times, there is no sample flow and the flow rate
shown is zero. Refer to
Note: To manually measure the flow rate through the analyzer, measure the flow rate at the analyzer drain
when the analyzer flushes the cell with water.
2. For standpipe installations, turn the shut-off valve at the Y-strainer outlet to set the flow rate
between 60 and 200 mL/min when the analyzer flushes the cell with water.
3. For pressure regulator installations, adjust the pressure regulator to set the flow rate between
60 and 200 mL/min when the analyzer flushes the cell with water.
3.10 Step 9: Install the latest software
Make sure that the SC Controller has the latest software installed. Use an SD card to install the latest
software on the SC Controller.
1. Go to the product page for the applicable SC Controller on http://hach.com.
2. Click the "Downloads" tab.
3. Scroll down to "Software/Firmware".
4. Click the link for the SC Controller software.
5. Save the files to an SD card.
6. Install the files on the SC Controller. Refer to the software installation instructions supplied with
the software files.
16 English
Table 4
on page 7 for descriptions of the measurement cycle steps.