RADWAG R Série Mode D'emploi page 97

Masquer les pouces Voir aussi pour R Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Cooperation with PUE 7.1, PUE 10 terminal
Format: NT CR LF
Response options:
ES CR LF
MASS FRAME
Response format:
NT
Stability marker
Zero marker
Range marker
Digit marker
Mass
Mass unit
Tare
Tare unit
Hidden
digit
- Number specifying quantity of hidden digits: [space] when no digits are hidden,
quantity
Balance status
- Number specifying balance status: [0] weighing, [1] pending adjustment start, [2]
Countdown prior
- For status [1], given in seconds as time interval prior to an automatic adjustment
to an automatic
adjustment
Example:
NT CR LF
- command sent form a computer
N T _ ? _ _ 0 _ _ _ _ _ - 5 . 1 1 3 _ g _ _ _ _ _ _ _ 0 . 0 0 0 _ g _ _ _0_1_28CR
- command carried out, response: mass value and remaining data where: _ - space
- command not understood (format error)
- response: mass value in a basic measuring unit
- command
- [space] if stable, [?] if unstable
- [space] for any value but zero, [Z] for zero value
- marker informing on range within which mass value is comprised: [space] I
range balance, [2] II balance, [3] III balance
[zero] no digit, [1] one digit marker, [2] two digits marker, [3] three digits
marker, [4] four digits marker, [5] five digits marker
10 characters for net mass given in adjustment unit (with dot and a floating-
point "-" sign )
3 characters - left justification
9 characters with dot - right justification (when floating-point tare gets turned
off automatically then zero value is sent)
3 characters - left justification
[1] when 1 digit is hidden, [2] when 2 digits are hidden, [3] when 3 digits are
hidden
balance adjustment
start, the value ranges between 30 - 01.
For the remaining statuses the value is 00.
- 97 -
LF