RADWAG R Série Mode D'emploi page 87

Masquer les pouces Voir aussi pour R Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Give tare value
Format: OT CR LF
Response: OT_TARE CR LF - command carried out
Response format:
1
2
3
O
T
space
Tare
- 9 characters with right justification
Unit
- 3 characters with left justification
CAUTION!
Tare value is always given in calibration unit.
Set tare
Format: UT_TARE CR LF, where TARE – tare value
Response options:
UT_OK CR LF
- command carried out
UT_I CR LF
- command understood but not accessible at this moment
ES CR LF
- command not recognised (tare format incorrect)
CAUTION!
Use dot in tare format as decimal point.
Send stable measurement result in a basic measuring unit
Format: S CR LF
Response options:
S_A CR LF
- command understood and in progress
S_E CR LF
- time limit exceeded while waiting for stable measurement result
S_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2-3
4
stability
S
space
marker
An example:
S CR LF
- command send form a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR
where: _ - space
Immediately send measurement result in a basic measuring unit
Format: SI CR LF
Response options:
SI_I CR LF
- command understood but not accessible at this moment
MASS FRAME
- immediate response: mass value in a basic measuring unit
Response format:
1
2
3
4
stability
S I
space
marker
An example:
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
where: _ - space
Send stable measurement result in a current measuring unit
Format: SU CR LF
Response options:
SU_A CR LF
- command understood and in progress
4-12
13
tare
space
5
6
space
character mass
- command carried out, response: mass value in a basic
LF
measuring unit.
5
6
space
character mass
- 87 -
14
15
16
unit
7-15
16
space
7-15
16
space
- command carried out, immediate response: mass
value in a basic measuring unit
17
18
19
space
CR
LF
17
18
19
20
unit
CR
17
18
19
20
unit
CR LF
21
LF
21

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

As r2.plusPs r1Ps r2Ps r2.h

Table des Matières