Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set min checkweighing threshold
Format:
DH_XXXXX CR LF, where: _ - space, XXXXX – mass format
Response options:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set max checkweighing threshold
Format:
UH_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Format:
ODH CR LF
Response: DH_MASA CR LF - command carried out
Response format:
1
2
3
D
H
space
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Give value of max checkweighing threshold
Format:
OUH CR LF
Response: UH_MASA CR LF - command carried out
Response format:
1
2
3
U
H
space
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Set mass value of a single item (only for PARTS COUNTING)
Format:
SM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Response options:
SM_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. mode other than
SM_I CR LF
PARTS COUNTING)
ES CR LF
- command not recognised (mass format incorrect)
Set target mass value (e.g. for DOSING)
Format:
TV_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
TV_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. mode other than
TV_I CR LF
DOSING)
ES CR LF
- command not recognised (mass format incorrect)
Set reference mass value (e.g. for PERCENT WEIGHING)
Format:
RM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Response options:
RM_OK CR LF
- command carried out
- command understood but not accessible at this moment (e.g. mode other than
RM_I CR LF
PERCENT WEIGHING)
ES CR LF
- command not recognised (mass format incorrect)
4-12
13
mass
space
4-12
13
mass
space
14
15
16
unit
14
15
16
unit
- 89 -
17
18
space
CR
17
18
space
CR
19
LF
19
LF