Maintenance; Technical Data And Measurement Tolerances - VOLTCRAFT LCR-9063 Notice D'emploi

L-c-r-mètre numérique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.3 Measurement of Resistances (R)
Important!
Make absolutely sure that all circuits and circuit components, as
well as other measurement objects, are no-load.
Connect the measuring leads supplied to the measuring device:
the red measuring lead to the „+" socket (red) and the black
measuring lead to the „−" socket (black)). Set the measurement
function switch to a range in the yellow sector („R") and set the
slide switch „R - L/C" to „R". With open measuring leads or high-
value/broken resistances „1" (for overflow) is displayed. Next,
connect the measurement probes to the measurement object. To
maintain the highest possible resolution (and thus the smallest
possible error of measurement), the measurement range must be
adjusted to the expected resistance. If, for example, the color
code rating indicates a value of 120 Ω, the range „200 Ω „ should
be set.
Note!
When you are measuring resistance, check that the measurement
points you touch with the measuring probes when measuring are
free of dirt, oil, soldering material, or similar, which are likely to
result in errors in measurement values.
With resistances greater than approx. 2 M Ω it may occur that the
display needs a few moments to stabilize itself. If „1" is display-
ed, you have exceeded the measurement range or the measure-
ment section has been interrupted.
12

6. Maintenance

Changing the battery is described in Section 4.1. For cleaning the
device or the display window, use only a clean, dry, antistatic lint-
free cleaning rag.
Important!
Never use cleaning agents containing carbon, petrol, alcohol or
similar since these will attack the surface finish of the measuring
device. Also, the vapors are a health and explosion hazard

7. Technical Data and Measurement Tolerances

7.1 Technical Data
Display . . . . . . . . . . . . . . . . .: 3-
Max. measurement rate . . .: 2.5 measurements per second
Overflow display . . . . . . . . .: „1" when range is exceeded
Environmental conditions
during the measurement . .: 0 °C to +50 °C, 0-80 % rel. humidi-
in-store . . . . . . . . . . . . . . . .: −20 °C to +60 °C, 0-80 % rel. humi-
Temperature coefficient . . .: 0.5 ppm/°C in the range below
Battery . . . . . . . . . . . . . . . . .: 9-V-block battery, alkaline, Code:
1
/
-digit LCD up to 1999, with
2
automatic polarity indicator (with
electrolytic capacitors)
ty, non-condensing
dity, after removal of the battery
from the measuring device
+18 °C and above +28 °C
NEDA 1604 or JIS 006P or IEC 6F22
13

Publicité

Table des Matières
loading

Table des Matières