Fault
Cause
number
E101
Firmware CRC
error
E102 /
RAM error
Err102
Err 103
Flash memory
error
Err 104
Event logger error
Err 105
Analog signal con-
verter error
E106
Incorrect software
installed
E107
Power supply error Connect power supply
E108
4-20 output error
E109
Sensor failure
Err 110
Sensor interface
failure
Err 112
Zero calibration
failed
Err 113
Span calibration
failed
Err 115
Gas value is under
range
Err 117
A magnetic key is
stuck "on" for more
than 1 minute
Err 119
Multiple consecu-
tive invalid gas
readings occurred
Err 121
PIR LOCK ON with
relays set to ENGR
OFF.
Err 122
ICOM data error
Dräger Polytron
Remedy
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
according to specifications.
Remove power; check the
field wiring from the Poly-
tron 5x00 to the control
system; reapply power.
Remove power; reconnect
or replace sensor; reapply
power.
Remove power; reconnect
or replace sensor; reapply
power; if this persists, call
Dräger Service.
Calibrate sensor.
Calibrate sensor.
Perform zero calibration.
Check environment for
possible gases with nega-
tive cross-sensitivity to the
measured gas.
Have the instrument
checked by DrägerService.
Check 'diag scan'; refer to
PIR 7x00 Technical Hand-
book section "Faults,
Cause and Remedy" for
more information.
Configuration conflict for
proper safety operation.
Reconfigure relay ENGR
settings to resolve.
Sensor communications
error. Remove power;
reconnect or replace sen-
sor; reapply power; if this
error persists call Dräger
Service.
®
5000 Series
Fault
Cause
number
Err 124
Beam Block error
Err 131
Cannot set PIR
lock
Err 198
Memory error
E198
Boot-load CRC
error
Err 199
Factory calibration
required

10.2 Warnings

Warning
Cause
number
Info 300
Instrument set-
tings and sensor
settings are differ-
ent
Info 301
Sensor is warming
up
Info 302
Beam block warn-
ing
Troubleshooting
en
Remedy
Clean optical surfaces; if
necessary, make sure that
hydrophobic filter, process
cuvette or process adapter
are correctly mounted.
Refer to PIR 7x00 Techni-
cal Handbook, section
"Check the Measuring
Cuvette of the Gas instru-
ment and Clean if
Required" and section
"Faults, Cause and Rem-
edy" for more information.
A1 and A2 settings and set
point are not identical. Cor-
rect relay settings before
engaging PIR LOCK.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Remedy
To select instrument set-
tings, tap [DOWN]. To
select sensor settings tap
[UP]. After selection is
made, display will show
'walking dashes'. When
data transfer is complete,
display will show 'CHk
CONF CAL'. Tap [OK] and
instrument will display
'walking dashes' then
return to normal operation.
Wait for warm-up phase to
end.
Clean optical surfaces; if
necessary, make sure that
hydrophobic filter, process
cuvette or process adapter
are correctly mounted.
Refer to PIR 7x00 Techni-
cal Handbook, section
"Check the Measuring
Cuvette of the Gas instru-
ment and Clean if
Required" and section
"Faults, Cause and Rem-
edy" for more information.